MicroScan Pos Combo Panel Type 33 (PC33): Catalog number B1017-211 MicroScan Pos Combo 33 B1017-211 panels are designed
FDA device recall Z-1524-2015 · posted 2015-04-27 · Terminated
Recall details
- Recalling firm
- Beckman Coulter, Inc.
- Reason for recall
- Siemens investigation confirmed that a portion one lot of panel Pos Combo 33 lot was packaged without desiccants. The substrates on a panel packaged without a desiccant may deteriorate over time due to exposure to additional moisture in the panel pouch.
- Action taken
- Siemens sent an Urgent Medical Device Recall letter dated April 8, 2014, to all affected customers. Letters advised users of the recalled product with catalog number, Siemens Material Number and Lot number as well as the reason for recall. The risk to health was explained as well as the recommendation for discussing the content of the letter with the medical director regarding the need to review previous test results. Customers are asked to discontinue use and discard any remaining inventory of the PC33 panel 2014-09-12. All other lot numbers are not impacted. The attached confirmation receipt is to be completed and returned and the letter should remain with laboratory records. Questions should be directed to Siemens Customer Care Center. For further questions call (916) 374-3031.
- Root cause
- Process control
- Status
- Terminated
- Product code
- JWY
- Quantity affected
- 664 panel boxes
- Distribution
- Nationwide Distribution including Puerto Rico and to the states of : CT, GA, NC, OK, NY, SC, KY, TN, CA, WA, OR, NJ, OH, PA, FL, MI, TX, WI, MO and NH.
- Date initiated
- 2014-04-08
- Date posted
- 2015-04-27
- Date terminated
- 2015-04-27
- Location
- West Sacramento, CA