Microscan plus Negative Combo 3 Sold outside of US only. Product Usage: MicroScan Synergies plus Panels & MicroScan Rapi
FDA device recall Z-2124-2013 · posted 2013-09-11 · Terminated
Recall details
- Recalling firm
- Siemens Healthcare Diagnostics, Inc.
- Reason for recall
- Confirmed increase in false susceptible and false intermediate misreads for Imipenem and/or meropenum on Synergies plus Negative and rap ID/S plus Negative panels read on Walk Away System.
- Action taken
- Siemens sent an Urgent Field Safety Notice letters dated August 2013 to all US customers via over night delivery, receipt requested. Letters for foreign customers are being translated and will be sent as soon as possible. The letter identified the affected product, reason for field correction, risk to health problem and actions to be taken by customer. Customers were instructed to suppress all reporting of sensitive and intermediate results for imipenem and meropenem. Siemens also recommends that the customer consider the need to review previous test results, conduct patient follow-up, and/or repeat testing. Customer were requested to complete the attached Field Correction Effectiveness Check form and fax it to 302-631-8467. For questions contact your Siemens Technical Support Provider or Distributor.
- Root cause
- Process control
- Status
- Terminated
- Product code
- LON
- Quantity affected
- 57,600 panels in total
- Distribution
- Worldwide Distribution - USA Nationwide in the states of: AK, SC, IA, MO GA, WI, TX, KS, CA, IL, CO, AZ, SD, IN, MD, FL, MN and the countries of:Japan, France, Algeria, Germany, Austria, Romania, Russ
- Date initiated
- 2013-08-21
- Date posted
- 2013-09-11
- Date terminated
- 2014-05-15
- Location
- West Sacramento, CA
Monitor recalls programmatically
Recalls and clearances share one API, with change alerts when new recalls post.
Related 510(k) clearances
| Record | Firm | Date |
|---|---|---|
| MICROSCAN SYNERGIES PLUS GRAM NEGATIVE MIC/COMBO PANELS WITH IMIPENEM (0.25, 1-32 UG/ML) (K020246) | Dade Behring, Inc. | 2002-02-06 |
| MICROSCAN PLUS GRAM NEGATIVE MIC/COMBO PANELS WITH MEROPENEM (0.12-32 UG/ML) (K020249) | Dade Behring, Inc. | 2002-02-06 |