MicroScan Neg Multidrug Resistant MIC 1 REF C32368 UDI-DI code: 15099590720568 Intended Use- For use with MicroScan Drie
FDA device recall Z-1794-2026 · posted 2026-04-08 · Open, Classified
Recall details
- Recalling firm
- Beckman Coulter, Inc.
- Reason for recall
- Due to manufacturing issue (incorrectly processing) with in vitro multidrug resistant test.
- Action taken
- On 09/08/2025, the firm sent via mail/email an "URGENT FIELD SAFETY NOTICE" to customers informing them that as a result of an internal investigation Beckman Coulter confirmed that Ciprofloxacin (Cp) antimicrobial agent was incorrectly manufactured on MicroScan Neg Multidrug Resistant MIC 1 Lot 2026-05-08. Customers are instructed to: "Review Ciprofloxacin results generated from Lot 2026-05-08 to identify any abnormal or unexpected trends in both susceptible and resistance patterns, or unexpected change in MIC trends, as compared to historical results for that same species (e.g. a sudden, unexplained increase in susceptible or % susceptible results for a species after implementation of Lot 2026-05-08 versus the previous susceptible or % susceptible results for the same species prior to using Lot 2026-05-08). "Discontinue use of this lot and discard per your laboratory guidelines "The laboratory should retain the inventory of other MicroScan Neg Multidrug Resistant MIC 1 lots as they are not impacted by this issue. Note-Beckman Coulter is no longer distributing MicroScan Neg Multidrug Resistant MIC 1 Lot 2026-05-08 For questions - contact Customer Support Center from website: http://www.beckmancoulter.com
- Root cause
- Under Investigation by firm
- Status
- Open, Classified
- Product code
- LTT
- Quantity affected
- 3 boxes
- Distribution
- International distribution to the country of Poland.
- Date initiated
- 2025-09-08
- Date posted
- 2026-04-08
- Location
- West Sacramento, CA