Microlab AT, part numbers 135001, 135001OOC, 135002, 135003, 135005, 135006, 135006R, 135007, 135008, 135016, 135028, 13

FDA device recall Z-0514-06 · posted 2006-02-14 · Terminated

Recall details

Recalling firm
Hamilton Co
Reason for recall
Heat sinking on power boards may not operate as intended and the board may cease to operate.
Action taken
On 12/15/05, the firm initiated the recall and its notification was via letters explaining the reason for the recall and requesting the product be returned.
Root cause
Other
Status
Terminated
Product code
JTC
Quantity affected
1071 units
Distribution
Product was released for distribution to 132 consignees in US, and 17 International consignees
Date initiated
2005-12-13
Date posted
2006-02-14
Date terminated
2008-05-07
Location
Reno, NV