Microbiologics KWIK-QC SLIDES ( KWIK-QC Acid Fast Stain Slides and KWIK-QC Mycobacterium Slides). Slide preparations ser
FDA device recall Z-1021-2013 · posted 2013-03-28 · Terminated
Recall details
- Recalling firm
- Microbiologics Inc
- Reason for recall
- Microbiologics Inc is initiating a voluntary recall on products KWIK-QC Acid Fast Stain Slide, SL42-10 and KWIK-QC Mycobacterium, SL43-10. These products are being recalled because the control organisms are not properly fixated on the slide, which may cause the user to unintentionally contaminate the patient area of the slide by flooding the control well with stain. This contamination could pote
- Action taken
- Microbiologists an "Urgent Medical Device Recall" letter dated February 11, 2013. The Letter described the product and the reason for the recall. Advised consignees to "Do not Distribute" and "Return" the product. Urged consignees to "Notify" end users and to complete the attached "Recall Response Form". Contact the form at 320-229-7057 for questions regarding this notice.
- Root cause
- Device Design
- Status
- Terminated
- Product code
- LJG
- Quantity affected
- 396 boxes of 10 slides each.
- Distribution
- Worldwide Distribution-USA including the states of CA, IN, MS, NY, OH, PA, and SD, and the countries of AUSTRALIA, AUSTRIA, UNITED ARAB EMIRATES, MALAYSIA, TAIWAN, JAPAN, TURKEY, BELGIUM, COLOMBIA, CA
- Date initiated
- 2013-02-11
- Date posted
- 2013-03-28
- Date terminated
- 2016-03-10
- Location
- Saint Cloud, MN