Micro-Tech Informer Plus Silver Monitor, Model 81830. The firm name on the label is Stanley Security Solutions, Inc., Li
FDA device recall Z-1476-2012 · posted 2012-04-28 · Terminated
Recall details
- Recalling firm
- Stanley Security Solutions, Inc.
- Reason for recall
- Connection between the call cord and the receptacle may not make a correct connection resulting in a continuous or intermittent (false) signal to the nurse call system
- Action taken
- Stanley Healthcare Solutions, Inc. contacted affected customers via phone and email on December 13, 2011. A follow-up certified letter was sent on December 14, 2011. Customers were additionally sent and "Important Recall Notice letter on December 16, 2011 after errors in the original letters were detected. The letter identified the affected product, problem, actions to be taken, and included a reply form for the customer to complete and return. The recalling firm requested their customers to stop using the monitor and contact them for return of the unit. Contact information (phone, fax, & email) was included. For question contact Micro-Tech Customer Service at 1-800-206-1045.
- Root cause
- Component design/selection
- Status
- Terminated
- Product code
- KMI
- Quantity affected
- 824 units
- Distribution
- Worldwide Distribution - USA Nationwide including the states of CT, FL, KY, MN, NC, NE, NH, NJ, NY, PA, TX and the country of Canada (5 units).
- Date initiated
- 2011-12-13
- Date posted
- 2012-04-28
- Date terminated
- 2013-09-24
- Location
- Lincoln, NE