Atlas FDA

Micro Curved Attachment (MCA) Cutting Burr, DFU 18-0001. All lots are included in recall. Part Number MCA Intended for u

FDA device recall Z-0749-2014 · posted 2014-01-15 · Terminated

Recall details

Recalling firm
The Anspach Effort, Inc.
Reason for recall
The Anspach Effort, Inc. is recalling all lots of the Anspach Micro Curved Attachment (MCA) Cutting Burrs and Dissection Tools. The recall was initiated after the indications for use were found to be broader than the validation of the MCA Cutting Burrs and its directions for use did not clearly state the correct intended use of the device. After a labeling correction, Anspach Effort, Inc. decide
Action taken
Firm is performing a Medical Device Recall by sending customers a recall letter and requesting the return of Anspach Micro Curved Attachments and Dissection Tools currently on hand.
Root cause
Labeling Change Control
Status
Terminated
Product code
HBE
Quantity affected
163
Distribution
Worldwide Distribution: US (nationwide) and Internationally to: AU, BE, CA, CH, ES, FR, GB, IT, JP, NL, SE, and ZA.
Date initiated
2012-03-29
Date posted
2014-01-15
Date terminated
2015-03-11
Location
Palm Beach Gardens, FL