MHI-TM2000 Linear Accelerator System, Software Version 3.0.0 revision 13 or earlier; Proprietary and Trade Name: MHI-TM2
FDA device recall Z-1931-2013 · posted 2013-08-08 · Terminated
Recall details
- Recalling firm
- MITSUBISHI HEAVY INDUSTRIES, LTD., HIROSHIMA MACHINERY WORK
- Reason for recall
- Software anomaly: Mitsubishi Heavy Industries, Ltd. became aware of a potential safety issue associated with the VERO (TM) Linear Accelerator System. The VERO (TM) may save treatment delivery record into wrong location under certain conditions.
- Action taken
- The firm, Mitsubishi Heavy Industries, Ltd. (MHI), sent an "URGENT FIELD SAFETY NOTICE" letter dated June 21, 2013 to its customers. The letter described the product, problem and actions to be taken. The customers were instructed to do the following: If any minor failure occurs, please always press [ALARM RESET] button on operator console first; and if [EXIT] button is clicked after a minor failure occurs, please click [YES] on the pop up dialog to finish treatment completely. MHI is developing a correction for this issue. Tentatively planned availability date: End of October 2013. The correction will be free of charge and a MHI Customer Service representative will contact affected customers to schedule its installation as the correction becomes available. If you need any further information or support concerning this issue, please feel free to contact our Customer Service representative: VERO GmbH, Kapellenstrasse 12, 8566 Feldkirchen Germany, at +49-89-99-1568-44 or email: franz.gum@vero-sbrt.com.
- Root cause
- Software Design Change
- Status
- Terminated
- Product code
- IYE
- Quantity affected
- 14 (2 in US)
- Distribution
- Worldwide distribution: US (nationwide) including states of TX and FL; and countries of: Belgium, Italy, Japan and Korea.
- Date initiated
- 2013-06-21
- Date posted
- 2013-08-08
- Date terminated
- 2017-03-23
- Location
- HIROSHIMA, 4-CHOME KAN-ON-SHIN-MACHI, Japan
Monitor recalls programmatically
Recalls and clearances share one API, with change alerts when new recalls post.
Related 510(k) clearances
| Record | Firm | Date |
|---|---|---|
| MHI-TM2000 LINEAR ACCELERATOR SYSTEM VERO IMAVIS (K122450) | Mitsubishi Heavy Industries, Ltd. | 2012-10-03 |
| MHI-TM2000 LINEAR ACCELERATOR SYSTEM (K072047) | Mitsubishi Heavy Industries, Ltd. | 2007-08-29 |