Metzenbaum Scissors 7, Curved, Sterile The Metzenbaum Scissors 7, Curved Curved are single-use sterile surgical instrume
FDA device recall Z-0752-2017 · posted 2016-12-13 · Terminated
Recall details
- Recalling firm
- Integra York PA, Inc
- Reason for recall
- The Chevron seals of 6x10 Tyvek peel pouch used to package the sterile instruments may separate causing the packages to open and compromise sterility of the instrument.
- Action taken
- The firm, Integra, mailed an "URGENT MEDICAL DEVICE RECALL NOTICE" dated October 26, 2016 to customers affected by the recall to notify them of the issue. The notice described the product, problem and actions to be taken. The customers were instructed to immediately review their inventory to determine if they had any of the recalled product on hand and to discontinue distributing the product; review your customer traceability records for shipment of the product and forward a copy of this Recall Notice to any of your customers that purchased the affected lots; complete and return the Acknowledgement Form that accompanied the notice (even if you do not have any product(s)) by email to: FCA3@integralife.com or fax to: 1-609-750-4220. Should you have any questions regarding these instructions, please contact Customer Service at 1-866-854-8300.
- Root cause
- Process control
- Status
- Terminated
- Product code
- LRW
- Quantity affected
- 50 units
- Distribution
- US Distribution in NC and TX.
- Date initiated
- 2016-10-28
- Date posted
- 2016-12-13
- Date terminated
- 2017-04-17
- Location
- York, PA