Atlas FDA

Metal Tray, 178 x 135 x 37 mm, REF: 11580A; Metal Tray, 275 x 178 x 35 mm, REF: 11580B; Metal Tray, 373 x 178 x 35 mm, R

FDA device recall Z-1991-2023 · posted 2023-06-21 · Open, Classified

Recall details

Recalling firm
Karl Storz Endoscopy
Reason for recall
Metal sterilization trays used to encase and protect medical devices during sterilization have a new 510(k) with instructions for use that is for the first time specific to the tray models, that also includes information regarding the sterilization modality.
Action taken
On 4/6/23, correction notices were distributed to customers who were informed the following: Affected metal trays are intended only for use to encase and protect specific KARL STORZ reusable medical devices for sterilization in pre-vacuum steam sterilization cycles (132C for 4 minutes with a 20-minute dry time). When used in conjunction with an FDA-cleared sterilization wrap, sterility of the enclosed medical device is maintained until used. The system has been validated with devices with working channels greater-than-or-equal-to 0.2 mm and a maximum length of 240 mm. 1) Ensure that all relevant personnel in your organization review this letter, including the new Instructions for Use for the metal trays: following: https://spwebspace.karlstorz.com/sites/MetalTrays/SitePages/Home.aspx 2) If you have previously sold or transferred any of the affected metal trays to a third party, please forward this notice to all such third parties. 3) Complete and return the Acknowledgment Form to karlstorz5983@sedgwick.com Customers correction-related questions can contact karlstorz5983@sedgwick.com or 888-943-5128. Customers with device-related questions can call 800-421-0837
Root cause
No Marketing Application
Status
Open, Classified
Product code
KCT
Quantity affected
4,933
Distribution
Worldwide - US Nationwide distribution including in the states of IN, MS, MN, AZ, FL, IL, CA, AK, NY, VA, TX, PA, LA, ND, AS, MD, NH, SC, AR, GA, NJ, NC, WA, WI, SD, TN, KY, AL, OK, MO, MA, MI, CT, UT
Date initiated
2023-04-06
Date posted
2023-06-21
Location
El Segundo, CA

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
KARL STORZ Metal Sterilization Trays (K203198)Karl Storz Endoscopy America, Inc.2021-07-06