META-TAN LAG/COMPRESSION SCREW KIT 90MM/85MM, REF: 71642690

FDA device recall Z-1806-2025 · posted 2025-05-19 · Open, Classified

Recall details

Recalling firm
Smith & Nephew, Inc.
Reason for recall
Integrated Lag/Compression Screw Kit has a compression screw with hexagon socket that may be undersized, which causes the screw to be non-functional, which would require replacement with another adequate size screw, and exchange may exceed 30 minutes.
Action taken
On 3/18/2025, recall notices were mailed and emailed to customers who were asked to do the following: 1) Inspect your inventory and locate any affected devices and quarantine them. 2) If you are a sales rep, district office, or distributor, you must notify your customers of the field action and ensure that the required actions are complete. 3) Complete and return the notice response section via email to FieldActions@smith-nephew.com If you have any questions email the firm at the email address above.
Root cause
Process control
Status
Open, Classified
Product code
HSB
Quantity affected
75
Distribution
Worldwide - US Nationwide distribution in the states of FL, NH, TX, CA, OH, NY, AL, IA, MA, NV, MO, WA, LA, TN, NC, KY, NM, NJ, AZ and the country of ZA.
Date initiated
2025-03-18
Date posted
2025-05-19
Location
Memphis, TN