Merlin@home RF Remote Monitoring Transmitter, Model EX1150, with software versions 6.1M or 7.0. Usage:RF Remote Transmit

FDA device recall Z-0949-2015 · posted 2015-01-09 · Terminated

Recall details

Recalling firm
St Jude Medical Cardiac Rhythm Management Division
Reason for recall
Transmitter may initiate a software reset resulting in backup operation in some implanted St. Jude Medical Radio Frequency (RF) enabled Implantable Cardioverter Assura, Unify Assura, and Quadra Assura ICDs and Assurity and Allure Pacemakers.
Action taken
The firm, St. Jude Medical, sent an "Important Medical Device Correction" Dear Doctor Letter dated 12/18/14 to customers to inform them that the Merlin@home transmitters may initiate a software reset resulting in backup operation in some implanted St. Jude Medical Radio Frequency (RF) enabled Implantable Cardioverter Defibrillators (ICDs) and Pacemakers. The letter informs the customers of the problems identified and the actions to be taken. Customers with questions or concerns are instructed to contact their local St. Jude Medical representative or St. Jude Medical's Technical Services Department at 1-800-722-3774.
Root cause
Device Design
Status
Terminated
Product code
NVZ
Quantity affected
257,456 units
Distribution
Worldwide Distribution: US (nationwide) and countries of: EU, Mideast, Canada, Australia/New Zealand, and Asia-Pacific (including Japan).
Date initiated
2014-12-18
Date posted
2015-01-09
Date terminated
2016-03-30
Location
Sylmar, CA