Merit Medical , Custom Inflation Kit REF: K05-00761F K05-01717B K05-02080D K05-02090C K05-02221 K05-02338 K05-02470 K05-
FDA device recall Z-1126-2026 · posted 2026-01-16 · Open, Classified
Recall details
- Recalling firm
- Merit Medical Systems, Inc.
- Reason for recall
- Inflation device handle may detach from the syringe during procedure.
- Action taken
- On December 4, 2025 Merit Medical issued a "Urgent Medical Device Recall Notice" to affected consignees via E-Mail. Merit asked consignees to take the following actions: 1. Please read and understand the recall notification. 2. Should the handle detach during use and the balloon not completely deflate, pull the handle out of the inflation syringe, disconnect the inflation syringe from the catheter, and use an alternate syringe to generate vacuum and aspirate residual contrast 3. Ensure that applicable personnel within your organization are made aware of this field action. 3. If the product has been further distributed to other facilities, institutions, or manufacturers, please ensure this notice is immediately shared with them. 4. Please fill out, scan, and email the completed Customer Response Form to Customer Service at RESPONSE@merit.com within seven (7) calendar days
- Root cause
- Process control
- Status
- Open, Classified
- Product code
- PPD
- Quantity affected
- 18,897 units
- Distribution
- Worldwide - Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, N
- Date initiated
- 2025-12-04
- Date posted
- 2026-01-16
- Location
- South Jordan, UT