MEGA 8Fr. IAB, Intra-Aortic Balloon Catheter and Accessories, used to provide counter pulsation therapy in the aorta Pro
FDA device recall Z-2220-2023 · posted 2023-07-21 · Open, Classified
Recall details
- Recalling firm
- Datascope Corp.
- Reason for recall
- During IAB catheter insertion, the introducer dilator may fracture at the hub, leaving the body of the introducer dilator housed within the sheath. Risks to health include damage to the femoral artery, descending aorta, or embolization; surgical intervention and delay of IAB therapy may also result. Firm has received 10 complaints, including 3 serious adverse events and 1 patient death.
- Action taken
- Firm notified customers on June 23, 2023 via letter. Customers were made aware of the potential for the device to break during use and the risk of using the device. Customers were given alternate solutions to continue using the IAB Catheter with alternative sheath/dilator combinations. Customers may also return their affected product for a credit. The firm will continue to ship product to customers who choose to continue receiving it. However, product currently being shipped is potentially affected by the same issue. Datascope/Getinge has identified the issue with the introducer dilator and is working with the supplier towards a resolution. If you have any questions, please contact your Datascope/Getinge representative or call Datascope/Getinge Customer Service at 1-888-943-8872, options 2, Monday through Friday, between the hours of 8:00 a.m. and 6:00 p.m. (Eastern Time Zone).
- Root cause
- Under Investigation by firm
- Status
- Open, Classified
- Product code
- DSP
- Quantity affected
- 321,609 total kits
- Distribution
- Domestic distribution nationwide. International distribution worldwide.
- Date initiated
- 2023-06-23
- Date posted
- 2023-07-21
- Location
- Fairfield, NJ