MEGA 7.5Fr. 40cc Intra-Aortic Balloon with Accessories, Model Number 0684-00- 0295-01
FDA device recall Z-0311-2025 · posted 2024-11-05 · Open, Classified
Recall details
- Recalling firm
- Datascope Corp.
- Reason for recall
- There is discrepant labeling on the inner and outer packaging of the device.
- Action taken
- An URGENT MEDICAL DEVICE REMOVAL notification letter was sent to customers beginning 9/23/24. Actions to be taken by the user: Our records indicate that you have received one or more of the affected MEGA 7.5Fr 40cc IAB devices. Please examine your inventory immediately to determine if you have any impacted product. See pictures of the outer cartons for the potentially impacted product below. Please verify all labels, IFU, and IAB against the artwork on the outer carton. If the product, IFU, and labels all match the information for the MEGA 7.5Fr 40cc outer carton (circled green in the second picture below), the device can be used as indicated in the IFU and does not need to be returned. If the outer carton states that Lot 3000344298 is a 50cc size (circled red in the first picture below), then this is incorrect and you should return the affected device by contacting Getinge QRC on QRC.UKI@getinge.com. Please forward this information to all current and potential Datascope/Getinge IAB users within your hospital/facility. If you are a distributor who has shipped any affected products to customers, please forward this letter to their attention for appropriate action. Whether or not you have any impacted product, complete and sign the attached Response Form (Page 5) to acknowledge that you have received and understand this notification. Return the completed form to Datascope/Getinge by e-mailing a scanned copy or by faxing the form to your local Datascope/Getinge Representative or office. Actions to be taken by Getinge: Datascope/Getinge is initiating this Urgent Medical Device Removal to notify users of this IAB labeling issue. This voluntary removal only affects the products identified above that are manufactured for lot number 3000344298; no other products are affected. Email: QRC.UKI@getinge.com
- Root cause
- Error in labeling
- Status
- Open, Classified
- Product code
- DSP
- Quantity affected
- 160 units
- Distribution
- International distribution to the countries of Germany, Spain , Finland, Great Britain, Hong Kong, Ireland, Italy, Lithuania, Moldova, Mexico, Norway, New Zealand, Poland, Sweden, Singapore.
- Date initiated
- 2024-09-17
- Date posted
- 2024-11-05
- Location
- Fairfield, NJ
Monitor recalls programmatically
Recalls and clearances share one API, with change alerts when new recalls post.
Related 510(k) clearances
| Record | Firm | Date |
|---|---|---|
| MEGA 8FR. 50CC AND SENSATION PLUS 8FR. 50CC INTRA-AORTIC BALLOON CATHETERS AND ACCESSORIES (K133074) | Datascope Corp. | 2013-12-12 |
| SENSATION(TM) PLUS 8FR. 50CC INTRA-AORTIC BALLOON CATHETER AND ACCESSORIES (K112327) | Cardiac Assist, Maquet Cardiovascular, LLC | 2011-09-09 |
| DATASCOPE 7.5FR. IAB CATHETER AND ACCESSORIES (K041281) | Datascope Corp. | 2004-06-07 |