Medtronic Temporary Pacemaker, Model 5348. Single Chamber External Pulse Generator (EPG) Temporary Pacemaker (AAI/VVI).
FDA device recall Z-1661-2014 · posted 2014-05-24 · Terminated
Recall details
- Recalling firm
- Medtronic Inc. Cardiac Rhythm Disease Management
- Reason for recall
- Events related to a pacing rate outside of the intended setting, including events of sudden increased pacing rate up to the maximum setting of 180 pulses per minute (ppm), have been reported. This is caused by the development of high resistance on internal electrical connector contacts due to oxidation over time.
- Action taken
- Consignees were sent a Medtronic "Urgent Medical Device Correction" letter dated April 2014. The letter was addressed to Risk Manger, Health Care Professional. The letter included the Issue Description, Root Cause, Potentially Affected 5348 EPGs within the Serial Number Ranges, Malfunction Indications, and Recommended Actions. For questions related to EPG service policy or purchasing a replacement EPG, customers should contact your Medtronic representative. For technical questions, please call Instruments Technical Service at (1-800)638-1991. Medtronic asks that customers complete and sign the attached Customer Confirmation Certificate and fax it to Medtronic at 651-367-0612 to the attention of Customer Focused Quality or e-mail to RS.CFOFCA@Medtrnic.com.
- Root cause
- Device Design
- Status
- Terminated
- Product code
- LWP
- Quantity affected
- 16,337 (8,989 US; 7,348 OUS)
- Distribution
- Worldwide Distribution -- USA, including the territories of Guam and Puerto Rico; and, the countries of Albania, Algeria, Australia, Austria, Bahamas, Bangladesh, Belarus, Belgium, Benin, Bolivia, Bos
- Date initiated
- 2014-04-29
- Date posted
- 2014-05-24
- Date terminated
- 2014-10-08
- Location
- Saint Paul, MN