Atlas FDA

Medtronic SmartLink Software loaded on SmartSync Patient Connector, Model Number 24967 with the following software compo

FDA device recall Z-2127-2025 · posted 2025-07-17 · Open, Classified

Recall details

Recalling firm
Medtronic, Inc.
Reason for recall
A software update for Medtronic CareLink SmartSync Device Manager (SmartSync) is now available. The update eliminates the potential display of an erroneous electrical reset (pop-up) message upon interrogation with a SmartSync Device Manager. The erroneous message indicates an electrical reset has occurred. However, no reset has actually occurred and there are no parameter changes or loss of diagnostics. Prior to this software update, if an erroneous SmartSync message was displayed, confusion regarding the message could have caused a clinician to consider an unnecessary system revision. Medtronic has received 39 reports of this display error with two instances of unnecessary device explant through 30 April 2025.
Action taken
Medtronic issued an URGENT MEDICAL DEVICE CORRECTION NOTIFICATION to its consignees on 06/18/2025 via UPS 2nd day delivery. The notice announced that a software update was available for the Medtronic CareLink SmartSync Device Manager (SmartSync), explained the issue which the update was correcting, risk, and requested the following: Customer Actions: " Update SmartSync (to version 4.2.3 or higher). Refer to release notes provided by your local Medtronic representative for update instructions. If needed, Medtronic representatives are available to work with you to install or update the SmartSync application(s) on your tablet. See Appendix for instructions on how to verify the software update is complete. " Sign and return the enclosed confirmation form and share this notice with those who need to be aware within your organization or with any organization where SmartSync may be in use. Questions regarding this information should be directed to the local Medtronic Representative or Medtronic Programmer Technical Services at 800-638-1991, Option 1.
Root cause
Software design
Status
Open, Classified
Product code
KRG
Quantity affected
3909 units
Distribution
Worldwide
Date initiated
2025-06-17
Date posted
2025-07-17
Location
Mounds View, MN

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.