Medtronic Mo.Ma Ultra Proximal Cerebral Protection Device, Model Number MOM0130068X5

FDA device recall Z-0322-2024 · posted 2023-11-17 · Open, Classified

Recall details

Recalling firm
Medtronic Inc
Reason for recall
The Mo.Ma Ultra device manifold is labelled with 2 stickers, an inferior and superior sticker, one on each side of the manifold. These stickers label the common carotid artery (CCA) proximal balloon and external carotid artery (ECA) distal balloon inflation and deflation ports. Both sides of the complaint units manifold were labelled with the superior label leading to an incorrect identification of the proximal and distal inflation /deflation ports when viewing the devices from the inferior side.
Action taken
Medtronic issued an URGENT: MEDICAL DEVICE RECALL NOTIFICATION on 10/04/2023 via UPS. The notice explained the issue, potential risk, and requested the return of the device. For questions, consignees were directed to contact their Medtronic Field Sales Representative.
Root cause
Process control
Status
Open, Classified
Product code
NTE
Quantity affected
48 units
Distribution
Worldwide distribution - US Nationwide and the countries of Algeria, Armenia, France, Germany, Iraq, Italy, Japan, Netherlands, Poland, Romania, Spain, Sweden, Switzerland, United Kingdom.
Date initiated
2023-10-12
Date posted
2023-11-17
Location
Minneapolis, MN