Medtronic MiniMed Paradigm Insulin Infusion Pumps for the following Paradigm models: MMT-511, MMT-512, MMT-712, MMT-515,
FDA device recall Z-1007-2007 · posted 2007-07-07 · Terminated
Recall details
- Recalling firm
- Medtronic MiniMed
- Reason for recall
- Exposure to Magnetic Resonance Imaging (MRI) has resulted in damage to the component that monitors and controls movement of the motor in the MiniMed Paradigm insulin infusion pump. Although there were alarms as a result of the damage, some users cleared these alarms and continued using the pump. Under such conditions, the pump will significantly over-deliver; potentially causing severe hypoglycemi
- Action taken
- Firm mailed Medical Device Safety letters on April 24, 2007, to healthcare professionals, existing pump users and MRI facilities reiterating the existing warning in the pump User Guide to avoid exposing the pump to MRI (or similar high strength electromagnetic fields) and strengthening previous warning by specifically mentioning the potential for over-delivery and severe hypoglycemia. Firm is also including an insert with this information with any Paradigm infusion pumps shipped to new customers.
- Root cause
- Other
- Status
- Terminated
- Product code
- LZG
- Quantity affected
- 334,000 Paradigm insulin pumps (approx)
- Distribution
- Worldwide: USA, US Virgin Islands, Antigua, Australia, Austria, Bahamas, Bahrain, Bermuda, Brazil, Canada, Chile, China, Columbia, Czech Republic, Denmark, Dominican Republic, Egypt, France, Germany,
- Date initiated
- 2007-04-27
- Date posted
- 2007-07-07
- Date terminated
- 2012-09-19
- Location
- Northridge, CA
Monitor recalls programmatically
Recalls and clearances share one API, with change alerts when new recalls post.
Related 510(k) clearances
| Record | Firm | Date |
|---|---|---|
| MEDTRONIC MINIMED PARADIGM INSULIN PUMP, MODEL MMT-712E (K053177) | Medtronic Minimed | 2006-01-31 |
| MEDTRONIC MINIMED PARADIGM INSULIN PUMP, MODELS MMT-515 AND MMT-715 (K040676) | Medtronic Minimed | 2004-05-21 |
| MEDTRONIC MINIMED PARADIGM INSULIN PUMP, MODEL MMT-712 (K031390) | Minimed, Inc. | 2003-07-23 |
| MEDTRONIC MINIMED PARADIGM MODEL 512 INSULIN PUMP AND THE BD PARADIGM LINK GLUCOSE METER (K030531) | Minimed, Inc. | 2003-06-17 |
| MINIMED PARADIGM INSULIN PUMP, MODEL 511 (K001829) | Minimed, Inc. | 2000-08-18 |