Medtronic Maximo II VR D284VRC, for ventricular pacing and defibrillation.

FDA device recall Z-0666-2009 · posted 2009-01-28 · Terminated

Recall details

Recalling firm
Medtronic Inc. Cardiac Rhythm Managment
Reason for recall
Maximo II VR ICD, DR ICD and CRT-D inadvertently include the Capture Management algorithm (Atrial/Right Ventricular/Left Ventricular Capture management). The Capture management algorithm monitors daily pacing thresholds and may increase or decrease pacing amplitudes in response to patients' needs. Although the algorithm is operating correctly, it is not accessible to the clinician for reprogramm
Action taken
A Medtronic Inc. letter dated July 30, 2008 was sent to Physicians. The letter described the product and problem. The letter advised to contact Technical Services for specific questions regarding the care of patients or how to program the Maximo II devices. The letter instructed to collect and hold all non-implanted Maximo II devices and that a Medtronic Representative will contact the doctor to make arrangements to update the programmer 2090.
Root cause
Process design
Status
Terminated
Product code
LWS
Quantity affected
376
Distribution
No US distribution. OUS to include: AUSTRIA, DENMARK, FINLAND, GERMANY, ITALY, SPAIN, SWEDEN, UK, GREECE, HONG KONG, INDIA, JORDAN, and SAUDI ARABIA.
Date initiated
2008-07-25
Date posted
2009-01-28
Date terminated
2011-05-07
Location
Mounds View, MN