Medtronic, Kappa 600 KDR600,KDR601,KDR603,KDR606,KDR651,KDR653 The devices are implantable pulse generators (IPG) used t
FDA device recall Z-2120-2011 · posted 2011-04-29 · Terminated
Recall details
- Recalling firm
- Medtronic Inc. Cardiac Rhythm Disease Management
- Reason for recall
- A rare measurement lock-up condition that may (at a rate of approximately 1 in 18,000 devices) inappropriately trigger ERI/RRT in the Medtronic Dual Chamber Pacemakers: Kappa 600, 700, 800, 900, EnPulse, Adapta, Versa, Sensia. This is caused by a random lock-up of the measurement system hardware that may result in an incorrect battery voltage reading of zero. This issue does NOT impact battery lon
- Action taken
- The firm, Medtronic, sent an "Important: Medical Device Correction" letter dated April 2011 to its customers. The letter described the product, problem, and actions taken. Medtronic informed the customers that this issue does NOT impact battery longevity and does NOT require device explant. Currently, the device can be reset to normal operation by a Medtronic representative. Reset devices are no more likely to experience a recurrence of this issue. Medtronic is planning to release a programmer software update that will allow the clinician to reset the device. In addition, Medtronic has consolidated their product performance data with the vision of creating a primary reference tool for performance updates. If you have any questions, please contact your local Medtronic Representative or Medtronic Technical Services at 800-505-4636.
- Root cause
- Component design/selection
- Status
- Terminated
- Product code
- NVZ
- Quantity affected
- 1,989,180
- Distribution
- Worldwide distribution: USA (nationwide) including Guam, Puerto Rico and DC; and countries including: Algeria, Andorra, Anguilla, Argentina, Armenia, Aruba, Australia, Austria, Bahamas, Bangladesh, Be
- Date initiated
- 2011-04-04
- Date posted
- 2011-04-29
- Date terminated
- 2012-11-27
- Location
- Saint Paul, MN