Atlas FDA

Medtronic Integrated Power Console System, IP X1 REF 1898001 or REF EC300 Medtronic Xomed, Inc., 6743 Southpoint Drive N

FDA device recall Z-0259-2010 · posted 2009-11-13 · Terminated

Recall details

Recalling firm
Medtronic Xomed, Inc.
Reason for recall
An error code (Code 13) is displayed on some of the consoles when a Medtronic Stylus or EHS handpiece is connected to the IPC indicating that the handpiece has not been recognized by the console.
Action taken
A Product Correction Notification was issued to each consignee on June 15, 2009. Medtronic sales representatives will visit each facility to advise consignees of the issue and to install a software upgrade that will correct the handpiece recognition issue on the console(s).
Root cause
Software design
Status
Terminated
Product code
ERL
Quantity affected
493
Distribution
Worldwide Distribution
Date initiated
2009-07-17
Date posted
2009-11-13
Date terminated
2010-01-20
Location
Jacksonville, FL