Medtronic implantable pulse generator: Sphera DR MRI SureScan, Dual chamber rate responsive pacemaker (DDDR): (a) Model

FDA device recall Z-0810-2019 · posted 2019-02-14 · Open, Classified

Recall details

Recalling firm
Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)
Reason for recall
A subset of Medtronic dual chamber pacemakers distributed worldwide between 10 March 2017 and 7 January 2019 under the brand names Adapta, Versa, and Sensia when programmed to a dual chamber mode with atrial-sensing, may experience a circuit error that affects device functionality
Action taken
In the US, beginning 17-Jan-2019, Medtronic Field Representatives hand deliver an FCA Notification Letter to consignees including implanting and follow-up physicians. In consultation with an Independent Physician Quality Panel (IPQP), patient management recommendations were provided with the FCA Notification letter to ensure patient safety. Medtronic Field Representatives will hand deliver an additional Supplemental Letter with the Urgent Medical Device Recall letter to a subset of physicians with patients whose device has shown evidence of a pacing pause that is potentially related to this circuit error. Consignees will be asked to return all unused and unopened affected product to Medtronic.
Root cause
Component design/selection
Status
Open, Classified
Product code
NVZ
Quantity affected
5392 units
Distribution
Wordlwide
Date initiated
2019-01-17
Date posted
2019-02-14
Location
Mounds View, MN