Atlas FDA

Medtronic Grafton Plus Paste a) DBM S45001 GRAFTON 1CC PASTE, REF S45001; b) DBM S45005 GRAFTON 5CC PASTE, REF S45005; c

FDA device recall Z-1329-2023 · posted 2023-04-04 · Open, Classified

Recall details

Recalling firm
Medtronic Sofamor Danek USA Inc
Reason for recall
Potential for packaging non-conformances directly related to the dual-barrier, sterile pouch packaging system. Non-conforming packaging may lead to a breach in the sterile barrier.
Action taken
Medtronic Sofamor Danek notified customers of the recall on about 02/08/2023, via letter. Customers were instructed to identify and quarantine any affected products, return any affected products, complete and return the Customer Confirmation Form, and provide the letter to all personnel that should be aware or where affected products have been transferred/distributed.
Root cause
Process control
Status
Open, Classified
Product code
MBP
Quantity affected
37721 units
Distribution
Worldwide distribution - US Nationwide and the countries of Australia, Austria, Brunei Darussalam, Canada, Chile, Colombia, Denmark, Dominican Republic, Ecuador, Finland, Ghana, Greece, Guatemala, Hon
Date initiated
2023-02-08
Date posted
2023-04-04
Location
Memphis, TN

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
GRAFTON PLUS DBM PASTE (K043048)Osteotech, Inc.2005-11-23