Medtronic Disposable Pressure Display Set, Models 61000, 61006, 62000, 62006. Sterilized by Ethylene Oxide. Non-pyrogeni
FDA device recall Z-0078-2012 · posted 2011-10-20 · Terminated
Recall details
- Recalling firm
- Medtronic Perfusion Systems
- Reason for recall
- Pressure Display Set (models 61000, 61006, 62000, 62006) DFU will be updated to clarify that the intended use of the disposables is only for cardiopulmonary bypass surgery (procedures up to 6 hours). Use beyond 6 hours may result in air leakage in the system resulting in a pressure reading that is not representative of the actual line pressure.
- Action taken
- Medtronic Perfusion Systems, sent an "Urgent Medical Device Notification letter dated September 9, 2011, via UPS 2-day delivery to all affected customers. The letter identified the affected devices, described the issues, and the actions needed to be taken. The letter indicates when approved, the updated Directions for Use for the Pressure Display Box and Pressure Display Sets will be sent to customers. Customers were asked to complete the Customer Confirmation Certificate and return to Medtronic per the instructions on the form. The letter states that if the recalled products have been further distributed, forward this letter immediately to all customers who received the affected products, and gather proof of notification. For further questions concerning this notification, contact Medtronic Cardiovascular Technical Support at 877-526-7890 or Medtronic Sales Representative.
- Root cause
- Device Design
- Status
- Terminated
- Product code
- DXS
- Quantity affected
- 193,590 units
- Distribution
- Worldwide Distribution - USA (nationwide) including Puerto Rico, and Virgin Islands (U.S.) and the countries of: Argentina, Australia, Austria, Belarus, Belgium, Bulgaria, Canada, Chile, China, Costa
- Date initiated
- 2011-09-09
- Date posted
- 2011-10-20
- Date terminated
- 2012-10-11
- Location
- Brooklyn Park, MN