Medtronic Carelink Monitor Model 2490C. The Medtronic CareLink Monitor Model 2490C is designed to automatically gather i
FDA device recall Z-2209-2010 · posted 2010-08-11 · Terminated
Recall details
- Recalling firm
- Medtronic Inc. Cardiac Rhythm Disease Managment
- Reason for recall
- Medtronic has identified that a subset of 2490C CareLink Monitors recently received an incorrect software update. Patients reported that their monitor was making a "ticking sound" and they were unable to transmit device data via CareLink. These monitors are now non-functional. No patient injuries have been reported as a result of this issue.
- Action taken
- Starting on 6/11/2010, Medtronic began sending Australian and Canadian consignees "Medical Device Recall" letters dated 6/10/2010. The letters describe the problem and product involved. The letters instructs consignees to order replacement monitors, notify affected patients, and instruct each patient to perform a manual transmission upon receipt of the new monitor. Medtronic will provide affected patients with a prepaid return kit to faciliate the return of the monitor to Medtronic. Australian consignees should contact Scott Dunlop at 61-7 3025-3547. Canadian consignees should contact Medtronic Patient Services at 1-800-268-5346.
- Root cause
- Software change control
- Status
- Terminated
- Product code
- NVZ
- Quantity affected
- 37 Foreign, 0 Domestic
- Distribution
- International Distribution Only: Australia and Canada.
- Date initiated
- 2010-06-11
- Date posted
- 2010-08-11
- Date terminated
- 2011-12-24
- Location
- Saint Paul, MN