Medtronic Affinity Pediatric Arterial Filter, model number S4014, sterilized using ethylene oxide. Recall Update 9/6/13:

FDA device recall Z-1966-2013 · posted 2013-08-14 · Terminated

Recall details

Recalling firm
Medtronic Inc. Cardiac Rhythm Disease Management
Reason for recall
Medtronic is recalling 148 Affinity Pediatric Arterial Filters from 2 manufacturing lots because a limited quantity of filters may have a small breach in the filter media that may allow unwanted particulate or gaseous emboli to pass through the filter and the outlet of the device, which could result in serious injury to a pediatric patient including neurological damage, or potential patient death.
Action taken
Customer communication was initiated verbally on June 14, 2013, to the 4 affected customers to quarantine any un-used units. An Urgent Medical Device Recall letter, dated June 26, 2013. was mailed to customers on June 28, 2013. The Urgent Medical Device letter referenced that the letter is in follow-up to the previous phone call. It identified the affected product, stated the issue with the 2 affected lot numbers and health risks, and asked that the affected devices be quarantined for return to Medtronic. If you have any questions please contact your Medtronic Sales representative or Lifeline Technical Services at 1-877-526-7890. 1-763-526-7890. Update: Customer communication to the newly identified customer was initiated verbally on August 20,2013, with the customer letter delivery beginning AUG 27, 2013. A Medtronic representative is available to facilitate the return of affected product.
Root cause
Employee error
Status
Terminated
Product code
DTM
Quantity affected
153 devices.
Distribution
Worldwide Distribution - US Distribution only to MS and IL, and the countries of Argentina, South Africa and Singapore.
Date initiated
2013-06-14
Date posted
2013-08-14
Date terminated
2014-03-07
Location
Saint Paul, MN