Medtronic, 10 Fr FlexCath Select Steerable Sheath 990065. The FlexCath Select Steerable Sheath (10 Fr Select) is used fo
FDA device recall Z-1276-2015 · posted 2015-03-17 · Terminated
Recall details
- Recalling firm
- Medtronic Inc. Cardiac Rhythm Disease Management
- Reason for recall
- Clinicians observed debris, appearing to originate from the hemostasis valve on the proximal end of the Steerable Sheath, outside of the patients bodies.
- Action taken
- The firm, Medtronic, Inc., sent an "URGENT MEDICAL DEVICE RECALL" letter dated February 2015 to its consignees/customers. Medtronic representatives, beginning Monday, Feb 23, 2015 hand delivered the letter to US consignees/customers.. The letter was addressed to Physician, Risk Manager and Health Care Professional. The letter described the product, problem and actions to be taken. The consignees/customers were instructed to immediately quarantine all unused product; return all affected product to Medtronic; contact Customer Service at 1-800-848-9300 to initiate product return and to complete and return the Customer Confirmation Certificate to via email to: RS.CFQFCA@Medtronic.com or fax to Medtronic at 651-367-0612 to the attention of Customer Focused Quality. In additions, consignees/customers should share this notification with others in your organization. Your Medtronic field representative may assist the customers with identifying suitable replacement product. If you have any questions, contact Customer Service at 1-800-848-9300.
- Root cause
- Nonconforming Material/Component
- Status
- Terminated
- Product code
- DYB
- Quantity affected
- 151 (4 US, 147 OUS)
- Distribution
- Worldwide Distribution: US (nationwide) in states of: NC and IA; and countries of: AUSTRIA, BELGIUM, DENMARK, FRANCE, GERMANY, NETHERLANDS, POLAND, SPAIN, and UNITED KINGDOM.
- Date initiated
- 2015-02-23
- Date posted
- 2015-03-17
- Date terminated
- 2015-10-07
- Location
- Saint Paul, MN