Medtronic 0-arm Imaging System, Product Catalog Number: B1-700-00027, Medtronic Navigation, Inc., Littleton, MA 01460 (s

FDA device recall Z-0294-2008 · posted 2007-11-27 · Terminated

Recall details

Recalling firm
Medtronic Navigation, Inc
Reason for recall
Reversed Instrument Position : a defect in the software version 3.0 causes an error in the position of surgical instruments and relative motion are reversed 180' on the StealthStation image. (If the StealthStation is not used, or if the patient orientation is prone or supine, the problem does not occur)
Action taken
Medtronic Navigation notified customers by letter on October 17, 2007. Medtronic Field service will perform upgrade (software version 3.0.1) at user site.
Root cause
Software change control
Status
Terminated
Product code
JZL
Quantity affected
7 units
Distribution
Worldwide: USA including states of CO, IN, NE, and MN and countries of Canada and Germany
Date initiated
2007-10-17
Date posted
2007-11-27
Date terminated
2007-11-27
Location
Louisville, CO