Atlas FDA

MEDLINE Urethral Catheterization Tray and Bag, Red Rubber Latex, 15 Fr, Medline Product Number/SKU (REF) DYND18350

FDA device recall Z-2482-2026 · posted 2026-06-17 · Open, Classified

Recall details

Recalling firm
Medline Industries, LP
Reason for recall
During an internal review, Medline identified representative samples did not meet applicable performance specifications. Identified issues include irregularities in catheter surface finish, instances where the catheter connector detached prematurely from the catheter funnel, and limited cases where the catheter outer diameter did not meet specification.
Action taken
Medline Industries issued two URGENT MEDICAL DEVICE RECALL notice on 5/4/2026 to its consignees, one to its catheter and catheter tray consignees, and the other to the kit consignees via USPS and email. Both notices explained the issue and potential risk. The catheter and catheter tray consignees were directed to respond to the notice and destroy the affected products. The kit consignees were directed to respond to the notice and apply additional labeling (over-labels) to affix to the kits which instruct the user to remove and destroy the affected component. Both notices direct distributors or those who further distributing the products to notify their customers. For questions, contact the Recall Department at 866-359-1704 or recalls@medline.com.
Root cause
Nonconforming Material/Component
Status
Open, Classified
Product code
OHR
Quantity affected
235658 units
Distribution
Worldwide distribution - US Nationwide and the countries of AE, BB, BM, BS, CA, CY, GU, NL, PA, QA, SA, VI.
Date initiated
2026-05-04
Date posted
2026-06-17
Location
Northfield, IL

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.