Atlas FDA

Medline procedure kits labeled as: 1) C SECTION, Pack Number DYNJ57622; 2) C SECTION PACK, Pack Number DYNJ48483B; 3) C

FDA device recall Z-1448-2025 · posted 2025-03-25 · Open, Classified

Recall details

Recalling firm
MEDLINE INDUSTRIES, LP - Northfield
Reason for recall
Medical convenience kits contain plastic syringes affected by the FDA Safety Alert issued on March 19, 2024. Leaks, breakage, and/other quality issues have been identified that may pose a risk to patient health.
Action taken
Medline Industries LP issued a recall notice to its consignees on 02/11/2025 via letter delivered via USPS and email. The notice explained the issue, provided a link to for information pertaining to the risk involved https://www.fda.gov/medical-devices/safety-communications/update-evaluating-plastic-syringes-made-china-potential-device-failures-fda-safety-communication) and request the syringes within the convenience be discarded, except for circumstances where no alternative syringes are readily available. If these syringes must be used, closely monitor for leaks, breakage, and any other problems, and report any quality issues to Medline and the FDA. Distributors were directed to notify their customers. This is an expansion of RES 94583
Root cause
Under Investigation by firm
Status
Open, Classified
Product code
OHM
Quantity affected
605 units
Distribution
Worldwide distribution - US Nationwide and the country of Canada.
Date initiated
2025-02-11
Date posted
2025-03-25
Location
Northfield, IL

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.