Medline Kits containing Codman Disposable Perforator 14 mm: 1. CRANI ACCESSORY PACK SKU DYNJ59270 2. CRANI PACK-LF SKU D

FDA device recall Z-1811-2025 · posted 2025-05-31 · Open, Classified

Recall details

Recalling firm
MEDLINE INDUSTRIES, LP - Northfield
Reason for recall
Affected kits contain recalled Integra Lifesciences Codman Disposable Perforators due to inadequate welds on specific lots. Perforators with inadequate welds may break or separate.
Action taken
On April 22, 2025, the firm notified customers of the recall through letters titled Immediate Action Required Voluntary Medical Device Recall. Customers were instructed to identify and quarantine all affected product on hand. Customers are to submit a response form to the recalling firm. Upon receipt of the response firm, Medline will provide the customer with over-labels to place on affected inventory, which will include instructions for staff to remove and discard the affected component prior to using the kit. Distributors should notify downstream customers/accounts of this recall. If you have any questions, contact the Recall Department at 866-359-1704 or recalls@medline.com.
Root cause
Nonconforming Material/Component
Status
Open, Classified
Product code
OJH
Quantity affected
1,853 total
Distribution
US distribution to CA, FL, IL, IN, KY, MA, MD, MN, NY, TX, VT, WA and WI. No OUS distribution.
Date initiated
2025-04-22
Date posted
2025-05-31
Location
Northfield, IL