Atlas FDA

Medline Kits containing B. Braun IV Administration Sets and Pump Administration Sets, labeled as: RR-LAP DONOR NEP ACCES

FDA device recall Z-1053-2026 · posted 2026-01-12 · Open, Classified

Recall details

Recalling firm
Medline Industries, LP
Reason for recall
Medline kits contain B. Braun IV Administration Sets and Pump Administration Sets which were recalled due to customer complaints identifying that the check valve components may become stuck in the open or closed positions.
Action taken
Medline issued an URGENT MEDICAL DEVICE CORRECTION notice to its consignees on 11/26/2025 via email and USPS first class mail. The notice explained the issue potential risk to the patient, and referred the consignee to the attached B Braun notice for further instructions. The B Braun notice provide "Interim Measures for Users" to provide for the continued use of the product. For questions, contact the Recall Department at 866-359-1704 or recalls@medline.com.
Root cause
Nonconforming Material/Component
Status
Open, Classified
Product code
RDE
Quantity affected
33 units
Distribution
US Nationwide distribution.
Date initiated
2025-11-26
Date posted
2026-01-12
Location
Northfield, IL

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.