Atlas FDA

Medline Hudson RCI Medium Concentration Oxygen Masks: 1041CE MASK,OXYGEN,UND CHIN,MED CO OUS ONLY; HUD1035 O2 MASK,MED C

FDA device recall Z-1035-2026 · posted 2026-01-08 · Open, Classified

Recall details

Recalling firm
Medline Industries, LP
Reason for recall
Firm received multiple complaints regarding tubing disconnecting from oxygen mask. If issue occurs during use, may lead to delay in patient care and patient may experience shortness of breath leading to hypoxia, which may require medical intervention to prevent further respiratory deterioration.
Action taken
On November 21, 2025, the firm notified customers via Urgent Medical Device Recall letter. Customers were instructed to destroy affected product on hand. No product is being returned to Medline. For kits, Medline instructed customers to request stickers to over-label the affected convenience kits with a warning label indicating to remove the affected component and replace it with product from supply.
Root cause
Under Investigation by firm
Status
Open, Classified
Product code
BYG
Quantity affected
1,611,081 eaches
Distribution
Worldwide distribution - US Nationwide and the countries of Australia, Bahamas, Canada, China, Dominican Republic, Guam, South Korea, Mexico, Panama, and Singapore.
Date initiated
2025-11-21
Date posted
2026-01-08
Location
Northfield, IL

Monitor recalls programmatically