Medical Surgical Bed, Model 3005S3EX Medical Surgical Bed The S3" MedSurg Bed, Model 3005 is intended to support and tra
FDA device recall Z-0629-2013 · posted 2013-01-03 · Terminated
Recall details
- Recalling firm
- Stryker Medical Division of Stryker Corporation
- Reason for recall
- Stryker Medical Model 3002S3EX and Model 3005S3EX beds built between March 15, 2012 and May 23, 2012. may have been built with brake rings that do not meet specifications, negatively impacting the holding force of the brakes.
- Action taken
- Stryker sent an Urgent Medical Device Correction letterdated December 18, 2012, to all affected customer. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to locate the affected product and ensure that the brakes are holding properly. Consignees were directed to contact Stryker (1-800-327-0770, option 2) to make arrangements to replace all brake rings and continue to routinely inspect brake functionality. Customers were asked to Fax (269-488-8691) the enclosed acknowledgment and Stryker field service will contact customers to schedule service. Consignees were directed to forward a copy of this notice to the new users and advise us of their new location. For questions regarding this recall call 269-488-8691.
- Root cause
- Process control
- Status
- Terminated
- Product code
- FNL
- Quantity affected
- 1004
- Distribution
- Worldwide Distribution - Nationwide and Canada
- Date initiated
- 2012-12-18
- Date posted
- 2013-01-03
- Date terminated
- 2013-08-30
- Location
- Portage, MI