Medical Gas Yoke Assembly. Generic 4-Cylinder Manifold Cart Model #8070. The device is a medical gas yoke assembly which
FDA device recall Z-0059-2010 · posted 2009-10-15 · Terminated
Recall details
- Recalling firm
- Western - a Scott Fetzer Company
- Reason for recall
- The yoke pins of the attached medical gas yoke assemblies could become displaced to such a degree that the yoke pin index safety system could become compromised and allow a post valve of one gas type to become interchangeable with a yoke connection of another gas type.
- Action taken
- Western Enterprises issued an "Urgent-Device Recall" notification dated August 20, 2009 informing Consignees of the affected product. Consignees are to contact the firm to request backup plates and installation instructions or request an RGA and shipping instructions for return of the manifold block assembly for modification. For further information, contact Western at 1-800-783-7890.
- Root cause
- Component design/selection
- Status
- Terminated
- Product code
- CAN
- Quantity affected
- 18 cart assemblies containing 72 yoke assemblies (36 Oxygen yokes and 36 Nitrous Oxide yokes)
- Distribution
- Nationwide Distribution -- AL, AR, CA, CO, FL, GA, HI, IL, IN, MA, OH, OR, PA, TX, WA and WI.
- Date initiated
- 2009-08-10
- Date posted
- 2009-10-15
- Date terminated
- 2012-04-25
- Location
- Westlake, OH