Atlas FDA

Medical Gas Yoke Assembly. Airgas Portable 2-Cylinder Manifold Cart Model #AGI-2237. The device is a medical gas yoke as

FDA device recall Z-0061-2010 · posted 2009-10-15 · Terminated

Recall details

Recalling firm
Western - a Scott Fetzer Company
Reason for recall
The yoke pins of the attached medical gas yoke assemblies could become displaced to such a degree that the yoke pin index safety system could become compromised and allow a post valve of one gas type to become interchangeable with a yoke connection of another gas type.
Action taken
Western Enterprises issued an "Urgent-Device Recall" notification dated August 20, 2009 informing Consignees of the affected product. Consignees are to contact the firm to request backup plates and installation instructions or request an RGA and shipping instructions for return of the manifold block assembly for modification. For further information, contact Western at 1-800-783-7890.
Root cause
Component design/selection
Status
Terminated
Product code
CAN
Quantity affected
42 cart assemblies containing 84 yoke assemblies (42 Oxygen yokes and 42 Nitrous Oxide yokes)
Distribution
Nationwide Distribution -- AL, AR, CA, CO, FL, GA, HI, IL, IN, MA, OH, OR, PA, TX, WA and WI.
Date initiated
2009-08-10
Date posted
2009-10-15
Date terminated
2012-04-25
Location
Westlake, OH