medex TranStar DPT w/Stopcock & 3cc Flush 10/EA, Product Code REF MX950SCFT

FDA device recall Z-0193-2025 · posted 2024-10-23 · Open, Classified

Recall details

Recalling firm
Smiths Medical ASD, Inc.
Reason for recall
Due to a manufacturing defect, users may experience inaccurate pressure monitoring, an interruption in pressure monitoring, or an inability to zero the device due to either a leak within the TranStar Disposable Pressure Transducer or from a defect within the TranStar Disposable Pressure Transducer chip.
Action taken
Smiths medical issued an URGENT MEDICAL DEVICE REMOVAL notice to its consignees on 09/24/2024 via traceable mailing. The notice explained the issue, potential risk, and requested the destruction of all affected product. Distributors were directed to forward the notice to their customers.
Root cause
Process control
Status
Open, Classified
Product code
DPT
Quantity affected
10 units
Distribution
Worldwide distribution.
Date initiated
2024-09-24
Date posted
2024-10-23
Location
Minneapolis, MN