Atlas FDA

Medefil Heparin I.V. Flush Syringe 1 unit/mL; a polypropylene luer locking syringe containing Heparin Lock Flush Solutio

FDA device recall Z-1545-2008 · posted 2008-05-09 · Terminated

Recall details

Recalling firm
Medefil Incorporated
Reason for recall
The heparin lock flush solution was manufactured from contaminated Heparin Sodium USP Active Pharmaceutical Ingredient (API). The FDA has received reports of serious injuries and/or deaths in patients administered finished heparin injectable products manufactured from heparin API containing this contaminant.
Action taken
Medefil telephoned the direct accounts on 3/20/08, e-mailed them a copy of the 3/20/08 recall letter on 3/21/08, and sent the accounts a hard copy of the recall letter on 3/24/08. The accounts were informed of Medefil's recall of 19 lots of Heparin I.V. Flush Syringes produced from the active pharmaceutical ingredient (API) Heparin Sodium, USP, lot 1035-0773, that has been recalled by the manufacturer, Scientific Protein Laboratories, due to the presence of a contaminant in the recalled lot of API. The accounts were instructed to stop use of the 19 lots of Heparin I.V. Flush Syringes immediately, place them under quarantine, and contact all customers to whom they further distributed the product, telling them to stop use of the product and return the product to Medefil through their supplier. The accounts were requested to complete and return to Medefil via fax the enclosed response form, indicating the quantity on hand of each recalled lot. Any questions were directed to Pradeep Aggarwal, President at 630-682-4600.
Root cause
Material/Component Contamination
Status
Terminated
Product code
NZW
Quantity affected
108,660 syringes
Distribution
Class II Recall - Nationwide Distribution --- including states of Arizona, Florida, Illinois, Massachusetts, Pennsylvania, and Texas.
Date initiated
2008-03-20
Date posted
2008-05-09
Date terminated
2010-04-15
Location
Glendale Heights, IL

Monitor recalls programmatically