Medcare Products Max-500 Stand, Total Support, patient stand, model #400006. The responsible firm name listed in the Ope
FDA device recall Z-0370-2014 · posted 2013-11-20 · Terminated
Recall details
- Recalling firm
- Ergosafe Products LLC DBA Prism Medical Services USA
- Reason for recall
- The lift might fail with the labeled 500-lb. load rating.
- Action taken
- The firm, Medcare Products, sent an "URGENT FIELD CORRECTIVE ACTIONS" letter dated October 21, 2013 via certified mail on October 30, 2013 to its customers. The letter described the product, problem and action to be taken. The customers were instructed to peel off the boom stickers and serial number label and replace with the new ones supplied; review the revised Manual and dispose of the original; and complete the action report form once the products have been relabeled. Please refer to your Operations Manual or visit our website at www.medcarelifts.com to view the manual online. If we can answer any questions that you may have regarding the testing standards or the relabeling of the products, please do not hesitate to contact us at 1-800-695-4479.
- Root cause
- Device Design
- Status
- Terminated
- Product code
- FSA
- Quantity affected
- 97 units
- Distribution
- Nationwide distribution.
- Date initiated
- 2013-10-30
- Date posted
- 2013-11-20
- Date terminated
- 2017-05-17
- Location
- Maryland Heights, MO
Monitor recalls programmatically
Recalls and clearances share one API, with change alerts when new recalls post.