Atlas FDA

Medacta Quadra Trial Broach size 0 part # 01.10.10.045, for use with the Medacta Total Hip Prosthesis System. Product Us

FDA device recall Z-0791-2013 · posted 2013-02-07 · Terminated

Recall details

Recalling firm
Medacta Usa Inc
Reason for recall
The tip of the Quadra Trial Broach size 0, broke during a surgery.
Action taken
Medacta sent a Urgent Safety Notification letter dated December 13, 2012 to all affected customers. The letter identified the recalled product, problem and actions to be taken. Customers were instructed to check inventory for affected product and contact QA@medacta.us.com for an RMA# to return items to Medacta USA. A new version of the recalled product will be provided to all affected customers. For questions call 805-437-7085 x26.
Root cause
Device Design
Status
Terminated
Product code
MEH
Quantity affected
26 pieces U.S.A. (226 pieces Worldwide)
Distribution
US Nationwide Distribution including the states of Arizona, California, Idaho, Illinois, North Carolina, Ohio, Utah, Washington and US Virgin Islands
Date initiated
2012-12-17
Date posted
2013-02-07
Date terminated
2013-02-07
Location
Camarillo, CA

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
MEDACTA TOTAL HIP PROSTHESIS SYSTEM - AMISTEM C & QUADRA C SHORT NECK (K103189)Medacta International S.A.2010-12-20
MEDACTA TOTAL HIP PROTHESIS SYSTEM-AMISTEN H , QUADRA H, QUADRA S (K093944)Medacta International S.A.2010-01-21
MEDACTA TOTAL HIP PROSTHESIS SYSTEM - QUADRA C (K083558)Medacta International S.A.2009-03-06
MEDACTA TOTAL HIP PROSTHESIS SYSTEM - QUADRA H AND R FEMORAL STEMS (K082792)Medacta International S.A.2008-12-29
TOTAL HIP PROSTHESIS SYSTEM QUADRA S + COCRMO FEMORAL BALL HEAD + APRICOT (K072857)Medacta International S.A.2008-02-04