McKesson Radiology-PACS. McKesson Radiology is a medical image and information application that is intended to receive,

FDA device recall Z-1617-2014 · posted 2014-05-13 · Terminated

Recall details

Recalling firm
Mckesson Medical Immaging
Reason for recall
High priority alert icon was not displayed in the Study List when there are no Stat studies present. However, the High priority alert icon is visible when studies marked as Stat priority exist in the Study List.
Action taken
On April 17, 2014, McKesson Medical Imaging Company initiated a process for applying a software solution to all McKesson Radiology customer systems that may be affected to prevent future occurrences of this issue. The systems with the potentially affected software versions will be corrected without removing the systems from their location.
Root cause
Software Design Change
Status
Terminated
Product code
LLZ
Quantity affected
45
Distribution
Worldwide Distribution-USA (nationwide) including the states of AL, CA, GA, ID, IL, IA, MD, MS, MO, NE, NV, NH, NY, NC, PA, TN, TX, VT, WA, WV, WY, and the countries of Canada and France.
Date initiated
2014-04-17
Date posted
2014-05-13
Date terminated
2016-02-02
Location
Richmond Canada, British Columbia