Atlas FDA

MC5057 12ft/3.6 meters Universal Cable. Made in USA. NON STERILE Rx Only. ConMed Linvatec, 11311 Concept Blvd, Largo, FL

FDA device recall Z-2007-2009 · posted 2009-10-01 · Terminated

Recall details

Recalling firm
Linvatec Corp. dba ConMed Linvatec
Reason for recall
Sufficiently worn/damaged Universal Cable with product number MC5057, when used with various ConMed Linvatec electric Handpieces, may cause the handpiece to self activate.
Action taken
ConMed Linvatec notified customers via letter, cautioning the users to inspect cords for signs of excessive wear or damage and urging them to discontinue use and return cables when appropriate. Direct questions to Linvatec Corporation by calling 1-800-237-0169.
Root cause
Component design/selection
Status
Terminated
Product code
HWE
Quantity affected
12,205 units
Distribution
Worldwide Distribution including United States.
Date initiated
2009-07-31
Date posted
2009-10-01
Date terminated
2011-07-11
Location
Largo, FL