Atlas FDA

MaxGuard Extension Set with Injection Site and 0.2 micron filter, Model No. ME3305

FDA device recall Z-1464-2017 · posted 2017-02-28 · Terminated

Recall details

Recalling firm
CareFusion 303, Inc.
Reason for recall
CareFusion is recalling the MaxGuard Extension Set with Injection Site and 0.2 micron filter because of reports of separation and/or leakages between the Y-connector and tubing.
Action taken
An Urgent Medical Device Recall Notification letter will be sent to customers on February 20, 2017 to inform them that CareFusion is recalling the MaxGuard Extension Set with Injection Site and 0.2 micron filter model number ME3305 Lot Number 16016790. Customers are informed that CareFusion has received reports of separation and/or leakages between theY-connector and tubing. Customers are informed that leakages from an extension set can cause delay of infusion, interruption of infusion, exposure to medication or hazardous infusates, or underinfusion. The letter informs the customers of the actions to be taken and for recall related questions to contact CareFusion Support Center at (888) 562-6018. Customers with adverse event reports are instructed to contact customer advocacy at (888) 812-3266.
Root cause
Process control
Status
Terminated
Product code
FPA
Quantity affected
1,200 units
Distribution
US Distribution including MA and MD
Date initiated
2017-02-20
Date posted
2017-02-28
Date terminated
2017-11-17
Location
San Diego, CA

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
INTRAVASCULAR ADMINISTRATION SET AND EXTENSION SET (K051499)Medegen Medical Manufacturing Services2005-06-22