Atlas FDA

MAS Omni Immune Controls

FDA device recall Z-1929-2014 · posted 2014-06-26 · Terminated

Recall details

Recalling firm
Microgenics Corporation
Reason for recall
The constituent, Inhibin A, was removed from MAS Omni IMMUNE and MAS Omni IMMUNE PRO controls; however the package insert was not updated to reflect the change.
Action taken
Urgent Medical Device Correction letters were sent on December 2, 2013 to all affected customers and distributors via trackable mail.
Root cause
Labeling Change Control
Status
Terminated
Product code
JJY
Quantity affected
3185
Distribution
Worldwide Distribution-USA (nationwide) and the countries of Australia, China, Germany, Hong Kong and Malaysia.
Date initiated
2013-12-02
Date posted
2014-06-26
Date terminated
2014-07-08
Location
Fremont, CA

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
THERMO SCIENTIFIC MAS OMNI-IMMUNE, MAS OMNI IMMUNE PRO (K110616)Microgenics Corp.2011-06-29