Mar Cor Purification Millenium HX Portable Reverse Osmosis System. Model 3025975. Product Usage: Water Purification Syst

FDA device recall Z-1213-2014 · posted 2014-03-14 · Terminated

Recall details

Recalling firm
Mar Cor Purification
Reason for recall
High inlet water pressure beyond specification and the solenoid valve does not meet application which could result in a delay in treatment.
Action taken
Mar Cor sent an Urgent Medical Device Correction letter dated March 5, 2014 to all affected customers. The letter identifies the affected devices, what the issue is and tells customers to contact Mar Cor to schedule correction. A response form was asked to be completed and returned. For questions contact Scott Nichols, Project Coordinator/Service Trainer at 1-800-633-3080.
Root cause
Component design/selection
Status
Terminated
Product code
FIP
Quantity affected
232 units
Distribution
USA Nationwide Distribution including Puerto Rico.
Date initiated
2013-06-26
Date posted
2014-03-14
Date terminated
2015-01-22
Location
Minneapolis, MN