Maquet VH-3010 Power Supply. Used to deliver power to the harvesting tool of the Vasoview Hemopro EVH System (HemoPro 2)

FDA device recall Z-1577-2025 · posted 2025-04-16 · Open, Classified

Recall details

Recalling firm
Maquet Cardiovascular, LLC
Reason for recall
Incorrect resistor utilized in the VH-3010 Power Supply, which may cause the jaws of the HemoPro 2 to not adequately heat up.
Action taken
An URGENT MEDICAL DEVICE REMOVAL notification dated March 2025 was mailed to consignees. The notification instructs consignees to examine their inventory for impacted serial numbers. For power supplies whose serial numbers are not listed in the provided Appendix A of recalled devices, devices can continue to be used. If power supplies with affected serial numbers are identified as being in stock, the associated devices should not be used and returned to Getinge. Returns are facilitated by contacting Getinge Customer Service at 888-800-2874 between the hours of 6:00 AM and 5:00 PM PST to request a return authorization number and shipping instructions. Consignees are to forward the provided notification to all users of the device within the facility and forwarded to any locations where devices were further distributed. Consignees with any questions about the recall are to contact their Getinge representative.
Root cause
Under Investigation by firm
Status
Open, Classified
Product code
GEI
Quantity affected
110 units
Distribution
Worldwide - US Nationwide distribution in the states of AL, CA, CO, DC, FL, GA, IA, ID, IL, IN, KY, MD, MI, MN, MO, NC, NH, NJ, NY, OH, PA, PR, TX, UT, VA, WI, WV and the countries of China, Japan, Ne
Date initiated
2025-03-10
Date posted
2025-04-16
Location
Wayne, NJ