Manual Biopsy Devices containing AIM system: Co-Axial Lung Biopsy Needle, 6 inch/15 cm, 22 gauge, Cat. #40-600, Material
FDA device recall Z-1445-04 · posted 2004-09-14 · Terminated
Recall details
- Recalling firm
- Boston Scientific Corporation
- Reason for recall
- Incomplete graduation on the needle guard, which would cause the graduations to be off by as much as 1 cm.
- Action taken
- Letters dated 8/9/04 hand delivered by Boston Scientific Sales Representatives. Sales Reps will visit each consignee and assist customers in locating affected units and will place a sticker label on each unit.
- Root cause
- Other
- Status
- Terminated
- Product code
- KNW
- Quantity affected
- 27 cartons x 5 units
- Distribution
- Products sold to 53 medical facilities nationwide and one consignee in Italy.
- Date initiated
- 2004-08-09
- Date posted
- 2004-09-14
- Date terminated
- 2005-10-26
- Location
- Natick, MA