Manual Biopsy Devices containing AIM system: Chiba Needle, 8 inch/20 cm, 22 gauge, Cat. #40-103, Material #M001401031 (1

FDA device recall Z-1443-04 · posted 2004-09-14 · Terminated

Recall details

Recalling firm
Boston Scientific Corporation
Reason for recall
Incomplete graduation on the needle guard, which would cause the graduations to be off by as much as 1 cm.
Action taken
Letters dated 8/9/04 hand delivered by Boston Scientific Sales Representatives. Sales Reps will visit each consignee and assist customers in locating affected units and will place a sticker label on each unit.
Root cause
Other
Status
Terminated
Product code
KNW
Quantity affected
48 cartons x 10 units
Distribution
Products sold to 53 medical facilities nationwide and one consignee in Italy.
Date initiated
2004-08-09
Date posted
2004-09-14
Date terminated
2005-10-26
Location
Natick, MA