FDA device recall Z-1638-2014 · posted 2014-05-20 · Terminated

Recall details

Recalling firm
Siemens Medical Solutions USA, Inc
Reason for recall
There is a potential and possible hazard to the user when using the MAMMOMAT Inspiration PC monitor at the control desk, in that the holder of the PC monitor can break causing an unstable monitor to fall causing possible serious injury.
Action taken
A safety advisory notice, dated April 25, 2014, was sent to direct accounts to notify them of the issue and provide instructions to avoid the potential risk of injury.
Root cause
Device Design
Status
Terminated
Product code
MUE
Quantity affected
5
Distribution
Distributed in the states of GA, FL, NY, NV, and MA.
Date initiated
2014-04-25
Date posted
2014-05-20
Date terminated
2014-12-15
Location
Malvern, PA